MACHINE TYPE:
High stability ice point reference to achieve the maximum possible precision for thermocouple calibrations.
APPLICATION SECTOR:
Pharmaceutical, chemical, cosmetic, food, dietary, medical devices.
FEATURES:
K140 Ice Point provides precise, stable, and reliable references at 0 °C for use in calibration laboratories. The patented design offers unparalleled accuracy (0.02 °C) and stability (0.01 °C) without long-term drift. The K140 model is ideal for calibrating any type of thermocouple or temperature sensor. The K140 offers 4 wells for sensor immersion. For applications where frequent connections or sensors need to be tested. It can support type T, J, K, B, E, N, R, and S thermocouples.
TECHNICAL SPECIFICATIONS
Dimensions (LxWxH) mm: 162x343x270
Power supply: 100-240V 50Hz
Manufacturer: Kaye
Model: K140
Year: 1997